ST103/106/112A Surge and HiPot Tester |
|
|
| >>Printable PDF |
| |
|
|
>>Contact Us |
|
|
How to Buy:
To purchase, contact your local Baker Sales Representative.
|
|
|
|
Overview
Proven reliability combined with rugged construction keeps this tester operating in the shop or in the field.
-
Greater current resolution for Polarization Index/Dielectric Absorption Tests and during HiPot Testing.
-
Total tester control from the front panel and easily accessible connectors.
-
All Leads Grounded selector position helps ensure operator safety.
-
Zero Start Interlock ensures operator safety and a low initial voltage to prevent insulation damage.
-
Versatility makes this tester an essential component for both corrective and predictive maintenance procedures. The capability to test all insulation systems of coils and windings makes this unit ideal for use on motors, generators, transformers, chokes, solenoids, and a variety of other coils.
-
Baker's exclusive QRR high-voltage component design.
-
Autoranging overcurrent Trip settings (low trip at 10µAmps), and visual Overcurrent Warning Indicator.
-
Leakage Current displayed on CRT gives quickest indication of faults, and provides a real-time, visual reference for controlling high-voltage testing.
-
Specially designed CRT display circuits are built for maximum reliability in the shop or in the field.
|
| |
| |
| |
| |
| |
| |
| |
| |
|
| |
Surge Test |
ST103A |
ST106A |
.ST112A |
Output Voltage |
3000V |
6000V |
12000V |
Max Output Current |
200A peak |
380A peak |
800A peak |
Max Pulse Energy |
.18 joules |
.72 joules |
2.88 joules |
| Minimum Test Inductance |
30-40µHenries |
30-40µHenries |
30-40µHenries |
| |
|
|
|
| DC HiPot |
|
|
|
| Max Output Voltage |
3000V |
6000V |
12000V |
| Max Output Current |
1000µA |
1000µA |
1000µA |
Overcurrent Trip |
10/100/1000 µA |
10/100/1000µA |
10/100/1000µA |
| Current Resolution |
1/10/100µA |
1/10/100µA |
1/10/100µA |
| |
|
|
|
| Physical Characteristics |
|
|
|
| Weight (lbs) |
42 lbs |
|
|
| Dimensions |
19" x 8" x 16" |
|
|
| |
|
|
|
| Power Requirements |
120VAC/118 Watts |
120VAC/118 Watts |
120VAC/333 Watts |
|
|
|
|
|
 |
Baker Instrument Company, an SKF Group Company
4812 McMurry Avenue
Fort Collins, CO 80525
970/282-1200 - 970/282-1010 FAX - 800/752-8272
©Copyright 2007, All Rights Reserved
|